9. You will then be asked to confirm the auto calibration values and the
instruments that the calibration cylinder will be used for. Be sure the auto
calibration values match the values listed on the calibration cylinder. The serial
number and gas combination of all selected instruments will appear
sequentially beneath the gas cylinder concentration. If you are bump testing a
standard 4-gas GX-2012 or GX-2012s, the screen will appear as shown below.
All channels will be bump tested using GAS 1.
Figure 38: Gas Order Confirmation Screen, Standard 4-Gas GX-2012