209 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations
9. Click OK. You will be asked if you want to calibrate the instruments. Click OK
to proceed. If you don’t want to proceed with the calibration, click Cancel to
return to the main program window.
10. The pump in each instrument will start and the program will indicate that air is
flowing. The AIR indicator on the program screen will be on. The CAL S LED
will begin to flash amber. Air will be applied to the instruments for 45 seconds,
then the PC Controller Program will perform a zero adjustment on all of the
channels.
Figure 150: Air Flowing
The PC Controller Program will analyze the zero adjustment results. If any
sensors failed the zero adjustment, the calibration will be aborted.