GX-2012 Operator’s Manual Using Maintenance Mode 123
11. Adjust the values for the warning and alarm settings using the procedure
in step 4 through step 7. When you press POWER ENTER after selecting
the alarm value, you will then be prompted to select the STEL value.
12. Use the AIRS or (SHIFT)T buttons to change the setting and press
POWER ENTER when finished to save the setting. You will then be
asked to enter a TWA value.
13. Use the AIRS or (SHIFT)T buttons to change the setting and press
POWER ENTER when finished to save the setting and return to the
Alarm Points Screen.
200
ppm
MAINTENANCE
STEL
CO
25
ppm
MAINTENANCE
TWA
CO
- - -
ppm
MAINTENANCE
ALARM--P
CO