Rki-instruments EAGLE 2 Manual Instrukcja Użytkownika Strona 1

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Czujniki Rki-instruments EAGLE 2 Manual. RKI Instruments EAGLE 2 Manual User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Operator’s Manual

www.rkiinstruments.comEAGLE 2Operator’s ManualPart Number: 71-0154RKRevision: SReleased: 7/31/14

Strona 2 - EAGLE 2 Operator’s Manual

Table of Contents EAGLE 2 Operator’s ManualStart Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 3 - Table of Contents

90 • Calibrating Using the Auto Calibration Method EAGLE 2 Operator’s Manual2. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is

Strona 4

EAGLE 2 Operator’s Manual Calibrating Using the Auto Calibration Method • 915. To adjust the values on the screen, hold down the RANGE ▼ SHIFT button,

Strona 5

92 • Calibrating Using the Auto Calibration Method EAGLE 2 Operator’s Manual12. Press and release the POWER ENTER RESET button. The following screen a

Strona 6

EAGLE 2 Operator’s Manual Calibrating Using the Auto Calibration Method • 9317. Fill the gas collection bag by screwing the fixed flow regulator or di

Strona 7

94 • Calibrating Using the Single Calibration Method EAGLE 2 Operator’s ManualIf any of the sensors cannot be adjusted to the proper value, a screen d

Strona 8

EAGLE 2 Operator’s Manual Calibrating Using the Single Calibration Method • 952. Press and release the POWER ENTER RESET button. The following screen

Strona 9

96 • Calibrating Using the Single Calibration Method EAGLE 2 Operator’s ManualPerforming a Span Adjustment in Single Calibration1. Slide the tubing cl

Strona 10

EAGLE 2 Operator’s Manual Calibrating Using the Single Calibration Method • 97and press and release POWER ENTER RESET to return to the Calibration Mod

Strona 11 - Chapter 1: Introduction

98 • Calibrating Using the Single Calibration Method EAGLE 2 Operator’s Manual12. Turn the knob clockwise to stop the gas flow, clamp down the hose cl

Strona 12

EAGLE 2 Operator’s Manual Calibrating Using the Single Calibration Method • 99In the example above, the EAGLE 2 could have adjusted the reading as low

Strona 13 - Specifications

EAGLE 2 Operator’s Manual Overview • 1Chapter 1: IntroductionOverviewThis chapter briefly describes the EAGLE 2 gas monitor. This chapter also describ

Strona 14 - About this Manual

100 • Parts List EAGLE 2 Operator’s ManualParts ListTable 12: Sample Bag Calibration Kit Spare PartsPart Number Description06-1248RK-03 Calibration ki

Strona 15

EAGLE 2 Operator’s Manual Overview • 101Appendix B: Setup ModeOverviewThis appendix describes the EAGLE 2 in Setup Mode. In Setup Mode, you can:• set

Strona 16 - Instrument Description

102 • Tips for Using Setup Mode EAGLE 2 Operator’s Manual• turn the confirmation alert feature on or off• turn the password feature on or off and set

Strona 17

EAGLE 2 Operator’s Manual Using Setup Mode • 103• Press the DISPLAY ADJUST NO button while in a screen where you are entering or updating parameters t

Strona 18

104 • Using Setup Mode EAGLE 2 Operator’s Manual5. If the unit prompts you for the password, enter it by using the AIR ▲ YES and RANGE ▼ SHIFT buttons

Strona 19 - S Sensors

EAGLE 2 Operator’s Manual Using Setup Mode • 1052. Press and release POWER ENTER RESET. The Set Date Format screen appears with the current setting fl

Strona 20 - Control Buttons

106 • Using Setup Mode EAGLE 2 Operator’s Manualconfiguration.CAUTION: Before changing the channel configuration, confirm that the correct sensors and

Strona 21 - I.S. Barrier PCB (not visble)

EAGLE 2 Operator’s Manual Using Setup Mode • 1074. Press and release POWER ENTER RESET. The available configuration options are shown. In the example

Strona 22 - Infrared Communications Port

108 • Using Setup Mode EAGLE 2 Operator’s Manual8. Use RANGE ▼ SHIFT to move the cursor next to the END menu item.NOTE: If you want to exit to the mai

Strona 23 - Battery Case & Batteries

EAGLE 2 Operator’s Manual Using Setup Mode • 109LEL is shown in terms of ppm. The available choices are on two screens.All of the gases except for the

Strona 24 - Flow System

2 • About the EAGLE 2 EAGLE 2 Operator’s Manual• Data logging functions (when used in Normal Mode)• Alarm trend data (when used in Normal Mode)• STEL

Strona 25 - Standard Accessories

110 • Using Setup Mode EAGLE 2 Operator’s ManualWARNING: Do not configure the catalytic sensor gas to hydrogen if you are monitoring for general hydro

Strona 26 - Sample Hose

EAGLE 2 Operator’s Manual Using Setup Mode • 111• 10000 ppm - - - *This selection is a user defined selection. The factory setting is 10000 ppm - -

Strona 27 - Battery Chargers

112 • Using Setup Mode EAGLE 2 Operator’s ManualNOTE: If you define a gas whose LEL is above 50,000 ppm, the %LEL reading in Measuring Mode will refle

Strona 28 - Optional Probes

EAGLE 2 Operator’s Manual Using Setup Mode • 113Mode.13. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired catalytic sensor voltage, 1.30 volts or

Strona 29 - Figure 12: Barhole Probe

114 • Using Setup Mode EAGLE 2 Operator’s Manual2. Press and release POWER ENTER RESET. The Catalytic Units screen appears with the current setting fl

Strona 30

EAGLE 2 Operator’s Manual Using Setup Mode • 115Updating the Alarm Point SettingsThis menu item allows you to update one or more alarm points (the rea

Strona 31 - 3:1 Dilution Fitting

116 • Using Setup Mode EAGLE 2 Operator’s Manual5. Press and release POWER ENTER RESET. The alarm point or alarm operation (oxygen only) will begin to

Strona 32 - Chapter 3: Operation

EAGLE 2 Operator’s Manual Using Setup Mode • 1171. From the main menu, place the cursor next to ALARM LATCHING.2. Press and release POWER ENTER RESET.

Strona 33

118 • Using Setup Mode EAGLE 2 Operator’s Manual3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.4. Press and release POWER ENTER RESE

Strona 34

EAGLE 2 Operator’s Manual Using Setup Mode • 1192. Press and release POWER ENTER RESET. The Adjust Auto Calibration Values Screen appears. The auto ca

Strona 35

EAGLE 2 Operator’s Manual Specifications • 3SpecificationsTable 1: Standard Sensor SpecificationsTable 2: EAGLE 2 SpecificationsCombustible Gas, Metha

Strona 36

120 • Using Setup Mode EAGLE 2 Operator’s Manual3. Use AIR ▲ YES and RANGE ▼ SHIFT to adjust the time to the desired setting.4. Press and release POWE

Strona 37 - OXY 20.9vol%

EAGLE 2 Operator’s Manual Using Setup Mode • 1212. Press and release POWER ENTER RESET. The Data Log Interval Screen appears.3. Use AIR ▲ YES or RANGE

Strona 38 - Continuous Operation

122 • Using Setup Mode EAGLE 2 Operator’s Manualshows the remaining log time and does not give you the opportunity to clear the logged data.1. From th

Strona 39 - Operation

EAGLE 2 Operator’s Manual Using Setup Mode • 1231. From the main menu, place the cursor next to CAL REMINDER.2. Press and release POWER ENTER RESET. T

Strona 40

124 • Using Setup Mode EAGLE 2 Operator’s Manual2. Press and release POWER ENTER RESET. The Calibration Past Due Action Screen appears. 3. Use AIR ▲ Y

Strona 41 - Combustible Gas Detection

EAGLE 2 Operator’s Manual Using Setup Mode • 125With LC/BH MODE SELECT set to LEAK CHECK ONLY, the Mode Select Screen appears when the unit is turned

Strona 42

126 • Using Setup Mode EAGLE 2 Operator’s Manual4. Press and release POWER ENTER RESET to save the setting and return to the main menu.Zero Follower S

Strona 43

EAGLE 2 Operator’s Manual Using Setup Mode • 127Turning the Password Function On or OffWith CHANGE PASSWORD set to ON, the EAGLE 2 prompts you for a p

Strona 44

128 • Using Setup Mode EAGLE 2 Operator’s Manualpassword is saved and you return to the main menu.Restoring the Default SettingsEach of the EAGLE 2 se

Strona 45 - Measuring Mode, Alarms

EAGLE 2 Operator’s Manual Using Setup Mode • 1294. Press and release AIR ▲ YES. A screen appears asking you to confirm that you want to restore the de

Strona 46 - Alarm Indications

4 • About this Manual EAGLE 2 Operator’s ManualAbout this ManualThe EAGLE 2 Operator’s Manual uses the following conventions for notes, cautions, and

Strona 47 - Sensor Failure

130 • Using Setup Mode EAGLE 2 Operator’s Manual1. From the main menu, place the cursor next to SPAN FACTOR.2. Press and release POWER ENTER RESET. Th

Strona 48 - Responding to Alarms

EAGLE 2 Operator’s Manual Using Setup Mode • 131Exiting Setup Mode1. From the main menu, place the cursor in front of NORMAL OPERATION at the bottom o

Strona 49 - Responding to Battery Alarms

132 • Overview EAGLE 2 Operator’s ManualAppendix C: Sub PCBsOverviewAn EAGLE 2 that has one or more of the standard four sensors, catalytic LEL, oxyge

Strona 50 - Display Mode

EAGLE 2 Operator’s Manual Channel Setup & Sub PCBs • 133retained in place with a screw/flat washer/lock washer. Any sub PCB can be installed in an

Strona 51 - Peak Screen

134 • Channel Setup & Sub PCBs EAGLE 2 Operator’s ManualCONFIGURE CHANNELS LCD screen below illustrates this situation.The “---” to the right of O

Strona 52 - Gas Display Screen

EAGLE 2 Operator’s Manual Channel Setup & Sub PCBs • 135• The currently configured gasThe catalytic sensor is always defined and configured even i

Strona 53

136 • Overview EAGLE 2 Operator’s ManualAppendix D: PID SensorsOverviewThe PID (photo ionization detector) sensors are used for applications where hig

Strona 54

EAGLE 2 Operator’s Manual Description • 137 Figure 25: Typical PID Sensor LocationPID Sensor & Sensor AdapterThe PID sensor is a cylindrical senso

Strona 55

138 • Start Up and Normal Operation EAGLE 2 Operator’s Manualuse the probe intended for PID use that does not have a particle filter installed and tha

Strona 56 - View Alarm Settings Screen

EAGLE 2 Operator’s Manual PID Relative Response Feature • 139state, the alternating numbers will be replaced by a gas reading. If the sensor doesn’t e

Strona 57

EAGLE 2 Operator’s Manual About this Manual • 5• Chapter 3 describes the operation of the EAGLE 2.• Chapter 4 describes Calibration Mode which allows

Strona 58 - Select User ID Screen

140 • PID Relative Response Feature EAGLE 2 Operator’s Manualisobutylene (IBL), you can select isopropyl alcohol (IPA) from a gas list accessible from

Strona 59

EAGLE 2 Operator’s Manual PID Relative Response Feature • 1412. With the PID Sensor Relative Response Screen displayed, press and release AIR ▲ YES. A

Strona 60 - Select Station ID Screen

142 • PID Relative Response Feature EAGLE 2 Operator’s ManualTable 15: High Range Relative Response Gas ListThe last choice in each list, PID, can be

Strona 61 - OPERATION

EAGLE 2 Operator’s Manual PID Calibration • 143PID CalibrationA PID channel can be calibrated using the auto calibration method or the single calibrat

Strona 62 - Data Logging Screen

144 • PID Calibration EAGLE 2 Operator’s Manual5. The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION.6. Move the cursor to

Strona 63 - Data Logging

EAGLE 2 Operator’s Manual PID Calibration • 1459. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate tha

Strona 64 - Data Logging Time

146 • PID Calibration EAGLE 2 Operator’s Manual13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button.14. Press and

Strona 65 - Chapter 4: Calibration Mode

EAGLE 2 Operator’s Manual PID Calibration • 147channel. The calibration gas value begins to flash.18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to a

Strona 66 - Equipment

148 • PID Calibration EAGLE 2 Operator’s Manualchanges when the unit is turned off and will load the previous set of auto calibration values when it i

Strona 67

EAGLE 2 Operator’s Manual PID Calibration • 14928. If all channels passed calibration, the following screen sequence occurs.If any of the sensors cann

Strona 68 - Calibrating Using the Auto

6 • Overview EAGLE 2 Operator’s ManualChapter 2: DescriptionOverviewThis chapter describes the EAGLE 2 instrument and accessories.Instrument Descripti

Strona 69

150 • PID Calibration EAGLE 2 Operator’s Manual31. If you want to change the PID channel’s calibration gas value, follow Step 5 - Step 24 above begin

Strona 70

EAGLE 2 Operator’s Manual PID Calibration • 15136. If the PID channel passed calibration, the following screen sequence occurs.If the PID channel cann

Strona 71

152 • PID Calibration EAGLE 2 Operator’s ManualCalibrating with a 5-Gas Cylinder1. See “Calibration Supplies and Equipment” on page 56 for a descripti

Strona 72

EAGLE 2 Operator’s Manual PID Calibration • 153If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Cali

Strona 73

154 • PID Calibration EAGLE 2 Operator’s Manual13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button.14. Press and

Strona 74

EAGLE 2 Operator’s Manual PID Calibration • 155channel. The calibration gas value begins to flash.18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to a

Strona 75

156 • PID Calibration EAGLE 2 Operator’s Manualchanges when the unit is turned off and will load the previous set of auto calibration values when it i

Strona 76

EAGLE 2 Operator’s Manual PID Calibration • 15728. If all channels passed calibration, the following screen sequence occurs.If any of the sensors cann

Strona 77

158 • PID Calibration EAGLE 2 Operator’s Manual29. If you want to change the PID channel’s calibration gas value, follow Step 5 - Step 24 above begin

Strona 78

EAGLE 2 Operator’s Manual PID Calibration • 15933. If the PID channel passed calibration, the following screen sequence occurs.If the PID channel cann

Strona 79

EAGLE 2 Operator’s Manual Instrument Description • 7Figure 2: Component Location, Side ViewFigure 3: Component Location, Front & BackTop CaseBotto

Strona 80 - Troubleshooting

160 • Maintenance EAGLE 2 Operator’s ManualMaintenanceThe PID sensor includes user serviceable parts. They are the lamp and the electrode stack. The f

Strona 81

EAGLE 2 Operator’s Manual Maintenance • 161• a small vial of aluminum oxide powder• 40 cotton swabs• 10 finger cotsPerform the following procedure to

Strona 82 - T

162 • Maintenance EAGLE 2 Operator’s Manual10. Hold the PID sensor steady on the working surface with one hand and using the other hand, locate the ta

Strona 83

EAGLE 2 Operator’s Manual Maintenance • 16316. Use this cotton swab to polish the PID lamp window. Use a circular motion, applying light pressure to c

Strona 84

164 • Maintenance EAGLE 2 Operator’s Manual20. Hold the electrode stack between the thumb and forefinger of one hand and place the window end of the l

Strona 85 - OFF ON CONTINUOUS

EAGLE 2 Operator’s Manual Maintenance • 16528. Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of t

Strona 86

166 • Maintenance EAGLE 2 Operator’s Manualusing the other hand, locate the tabs on the electrode stack removal tool and insert them into the slots on

Strona 87

EAGLE 2 Operator’s Manual Maintenance • 167Figure 32: Reinstalling the Electrode Lamp16. Twisting the lamp slightly during insertion will help to ensu

Strona 88

168 • Maintenance EAGLE 2 Operator’s Manual24. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on

Strona 89 - Charcoal Filter

EAGLE 2 Operator’s Manual Maintenance • 169NOTE: At this point it is recommended that the finger cots be used on the fingers handling the lamp. Finger

Strona 90 - Replacing a Sensor

8 • Instrument Description EAGLE 2 Operator’s ManualCaseThe EAGLE 2’s sturdy, high-impact plastic case is radio frequency (RF) resistant and is suitab

Strona 91 - Figure 22: Replacing a Sensor

170 • Maintenance EAGLE 2 Operator’s ManualFigure 34: Reinstalling Electrode Lamp16. Twisting the lamp slightly during insertion will help to ensure t

Strona 92

EAGLE 2 Operator’s Manual Maintenance • 171is resting on the bottom of the bottom case. If the main PCB is not seated properly, it may be damaged when

Strona 93

172 • Configuring the PID Gas in Setup Mode EAGLE 2 Operator’s Manual12. Insert the sensor adapter into the PID flow chamber and push it in until it b

Strona 94 - General Parts List

EAGLE 2 Operator’s Manual Configuring the PID Gas in Setup Mode • 1734. The “S” will then disappear and the following screen will appear for a few sec

Strona 95

174 • Configuring the PID Gas in Setup Mode EAGLE 2 Operator’s Manual8. Use RANGE ▼ SHIFT to move the cursor down the menu to the PID sensor.9. To cha

Strona 96 - Calibration Tubing

EAGLE 2 Operator’s Manual Configuring the PID Gas in Setup Mode • 17512. Use AIR ▲ YES and RANGE ▼ SHIFT to place the cursor next to the desired gas.

Strona 97 - Entering Calibration Mode

176 • Configuring the PID Gas in Setup Mode EAGLE 2 Operator’s Manual13. Press and release the POWER ENTER RESET button. The LCD will ask for confirma

Strona 98

EAGLE 2 Operator’s Manual Configuring the PID Gas in Setup Mode • 17719. A screen appears that asks if you want to save the changes you have made.NOTE

Strona 99

178 • Parts List EAGLE 2 Operator’s ManualParts ListTable 17: PID EAGLE 2 Spare PartsPart Number Description33-0560RK Electrode stack, 0 - 50 ppm, 2 p

Strona 100

EAGLE 2 Operator’s Manual Overview • 179Appendix E: ESM-01 Toxic SensorsOverviewThe ESM-01 sensors are used to monitor levels of a variety of toxic ga

Strona 101

EAGLE 2 Operator’s Manual Instrument Description • 9Oxygen SensorThe O2 sensor is a galvanic type of sensor. A membrane behind the openings on the sen

Strona 102

180 • Description EAGLE 2 Operator’s ManualSee “Replacing the ESM-01 Sensor or Changing Sensor Type” on page 199 for instructions to replace or change

Strona 103

EAGLE 2 Operator’s Manual Start Up and Normal Operation • 181ESM-01 Sub PCBThe ESM-01 sub PCB is a circuit board that is installed on the main PCB in

Strona 104 - Calibration Method

182 • ESM-01 Calibration EAGLE 2 Operator’s Manualdescribed below in “Calibrating with a 5-Gas Cylinder” on page 191, uses a multi-gas cylinder that c

Strona 105

EAGLE 2 Operator’s Manual ESM-01 Calibration • 1836. Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT button.7. Press an

Strona 106

184 • ESM-01 Calibration EAGLE 2 Operator’s Manual9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate

Strona 107

EAGLE 2 Operator’s Manual ESM-01 Calibration • 18513. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button.14. Press a

Strona 108

186 • ESM-01 Calibration EAGLE 2 Operator’s Manualchannel. The calibration gas value begins to flash.18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons t

Strona 109 - H2S CO

EAGLE 2 Operator’s Manual ESM-01 Calibration • 187changes when the unit is turned off and will load the previous set of auto calibration values when i

Strona 110 - Parts List

188 • ESM-01 Calibration EAGLE 2 Operator’s Manual28. If all channels passed calibration, the following screen sequence occurs.If any of the sensors c

Strona 111 - Appendix B: Setup Mode

EAGLE 2 Operator’s Manual ESM-01 Calibration • 18931. If you want to change the ESM-01 channel’s calibration gas value, follow Step 15 - Step 24 above

Strona 112 - Tips for Using Setup Mode

EAGLE 2 Operator’s ManualWARNINGRead and understand this instruction manual before operating instrument. Improper use of the gas monitor could result

Strona 113 - EAGLE 2 S

10 • Instrument Description EAGLE 2 Operator’s ManualControl ButtonsFour control buttons are located below the LCD. They are, from left to right, POWE

Strona 114 - Setting the Date Format

190 • ESM-01 Calibration EAGLE 2 Operator’s Manual36. If the ESM-01 channel passed calibration, the following screen sequence occurs.If the ESM-01 cha

Strona 115 - Configuring the Channels

EAGLE 2 Operator’s Manual ESM-01 Calibration • 191Calibrating with a 5-Gas CylinderThe calibration instructions below show a 5 channel instrument whic

Strona 116

192 • ESM-01 Calibration EAGLE 2 Operator’s Manual7. Press and release the POWER ENTER RESET button. The following screen appears.8. Press and release

Strona 117

EAGLE 2 Operator’s Manual ESM-01 Calibration • 193the probe are tightened firmly together to avoid leaks that can affect the calibration. See Figure 1

Strona 118

194 • ESM-01 Calibration EAGLE 2 Operator’s Manual16. Place the cursor next to the channel whose gas value you want to change using the AIR ▲ YES and

Strona 119

EAGLE 2 Operator’s Manual ESM-01 Calibration • 195If you select NO by pressing and releasing the DISPLAY ADJUST NO button, the changes you made will b

Strona 120

196 • ESM-01 Calibration EAGLE 2 Operator’s Manual28. If all channels passed calibration, the following screen sequence occurs.If any of the sensors c

Strona 121

EAGLE 2 Operator’s Manual ESM-01 Calibration • 19729. If you want to change the ESM-01 channel’s calibration gas value, follow Step 15 - Step 24 above

Strona 122 - UP/DOWN THEN ENTER

198 • ESM-01 Calibration EAGLE 2 Operator’s Manual33. If the ESM-01 channel passed calibration, the following screen sequence occurs.If the ESM-01 cha

Strona 123

EAGLE 2 Operator’s Manual Maintenance • 199MaintenanceThe SO2 and HCN ESM-01 sensors are the only ESM-01 sensors that include user serviceable parts.

Strona 124

EAGLE 2 Operator’s Manual Instrument Description • 11Printed Circuit Boards (PCBs)Figure 4: EAGLE 2 PCBsThe EAGLE 2’s PCBs analyze, record, control, s

Strona 125

200 • Maintenance EAGLE 2 Operator’s Manualthe ESM-01 chamber.14. Install the two sensor bracket screws.15. Confirm that the main PCB is seated in its

Strona 126

EAGLE 2 Operator’s Manual Maintenance • 201in the sensor body by a threaded collar on the connector end of the sensor. Unscrew the collar from the sen

Strona 127

202 • Parts List EAGLE 2 Operator’s Manualthe ESM-01 chamber.21. Install the two sensor bracket screws.22. Confirm that the main PCB is seated in its

Strona 128 - STATION ID’S

EAGLE 2 Operator’s Manual Parts List • 20381-0185RK-02 Calibration cylinder, 0.5 ppm PH3 in nitrogen, 58 liter81-0192RK-02 Calibration cylinder, 2 ppm

Strona 129

204 • Overview EAGLE 2 Operator’s ManualAppendix F: TC SensorsOverviewThe TC sensors are used to monitor high levels of combustible gas. This appendix

Strona 130 - ON POWERUP

EAGLE 2 Operator’s Manual Description • 205Figure 37: Typical TC Sensor LocationTC SensorThe TC sensor’s appearance is exactly the same as that of the

Strona 131

206 • Start Up and Normal Operation EAGLE 2 Operator’s ManualTC sensor cable connects to the sub PCB. The sub PCB plugs into the main PCB and is held

Strona 132

EAGLE 2 Operator’s Manual TC Calibration • 2071. Use the DISPLAY ADJUST NO button to enter Display Mode and scroll to the Catalytic (LEL) Sensor scree

Strona 133 - REMINDER

208 • TC Calibration EAGLE 2 Operator’s Manualappropriate for the TC detection range.If the EAGLE 2 is intended for use in a landfill, RKI Instruments

Strona 134 - Updating LC/BH Mode Setting

EAGLE 2 Operator’s Manual TC Calibration • 2097. Press and release the POWER ENTER RESET button. The following screen appears.8. Press and release the

Strona 135

12 • Instrument Description EAGLE 2 Operator’s ManualFigure 5: Main PCBThe main PCB is located on the right side of the bottom case. It slides into gu

Strona 136 - CONFIRMATION ALERT

210 • TC Calibration EAGLE 2 Operator’s Manualcomplete with internal O-ring and membrane and that the two halves of the probe are tightened firmly tog

Strona 137

EAGLE 2 Operator’s Manual TC Calibration • 21115. To adjust the values on the screen, press and hold the RANGE ▼ SHIFT button, then press the DISPLAY

Strona 138 - CONFIGURATION?

212 • TC Calibration EAGLE 2 Operator’s Manual22. Press and release the POWER ENTER RESET button. The following screen appears.23. If you select YES b

Strona 139

EAGLE 2 Operator’s Manual TC Calibration • 21327. Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the pr

Strona 140 - Updating the Language Setting

214 • TC Calibration EAGLE 2 Operator’s Manual31. If you want to change the TC channel’s calibration gas value, follow Step 15 - Step 24 above beginni

Strona 141 - CHANGES IN MEMORY?

EAGLE 2 Operator’s Manual TC Calibration • 21536. If the TC channel passed calibration, the following screen sequence occurs.If the TC channel cannot

Strona 142 - Sub PCB Positions

216 • Maintenance EAGLE 2 Operator’s ManualMaintenanceThe TC sensor does not include any user serviceable parts. This section includes a procedure for

Strona 143 - CH4: OP1 (PID)

EAGLE 2 Operator’s Manual Configuring the TC Gas in Setup Mode • 217properly, it may be damaged when the top case is reinstalled.16. Make sure that th

Strona 144 - Sub PCBs and CONFIGURE GASES

218 • Configuring the TC Gas in Setup Mode EAGLE 2 Operator’s Manual4. The “S” will then disappear and the following screen will appear for a few sec

Strona 145 - IS NOT POSSIBLE

EAGLE 2 Operator’s Manual Configuring the TC Gas in Setup Mode • 2198. Use RANGE ▼ SHIFT to move the cursor down the menu to the TC sensor.9. To chang

Strona 146 - Description

EAGLE 2 Operator’s Manual Instrument Description • 13infrared port or an IrDA/USB cable connected to a USB port can be used to download data saved by

Strona 147 - PID Probe

220 • Configuring the TC Gas in Setup Mode EAGLE 2 Operator’s ManualWARNING: Do not use the user defined gas configuration without consulting RKI Inst

Strona 148 - Start Up and Normal Operation

EAGLE 2 Operator’s Manual Configuring the TC Gas in Setup Mode • 22121. Press and release POWER ENTER RESET.22. A screen appears that asks if you want

Strona 149 -

222 • Parts List EAGLE 2 Operator’s ManualParts ListTable 21: TC EAGLE 2 Parts ListPart Number Description47-5010RK TC/LEL sensor cableTE-7568 TC sens

Strona 150 - FOR PID SENSOR

EAGLE 2 Operator’s Manual Overview • 223Appendix G: Infrared Carbon Dioxide SensorsOverviewThe infrared CO2 sensors are used to monitor levels of carb

Strona 151

224 • Description EAGLE 2 Operator’s Manual Figure 38: Typical Infrared CO2 Sensor LocationThe infrared CO2 sensor is installed in a single sensor flo

Strona 152

EAGLE 2 Operator’s Manual Description • 225one of the 3 sub PCB positions when an infrared CO2 sensor is used with the EAGLE 2. The infrared CO2 senso

Strona 153 - PID Calibration

226 • Start Up and Normal Operation EAGLE 2 Operator’s ManualStart Up and Normal OperationFor instructions to startup and use an EAGLE 2 that includes

Strona 154

EAGLE 2 Operator’s Manual Infrared CO2 Calibration • 2275. Put the caps back on the scrubber.Infrared CO2 CalibrationAn infrared CO2 channel can be ca

Strona 155

228 • Infrared CO2 Calibration EAGLE 2 Operator’s Manual5. The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION.6. Move the c

Strona 156

EAGLE 2 Operator’s Manual Infrared CO2 Calibration • 22911. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then in

Strona 157

14 • Instrument Description EAGLE 2 Operator’s ManualFlow SystemThe EAGLE 2 flow system consists of the inlet fitting, hydrophobic filter, pump, inter

Strona 158

230 • Infrared CO2 Calibration EAGLE 2 Operator’s Manual17. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button.18. P

Strona 159

EAGLE 2 Operator’s Manual Infrared CO2 Calibration • 231channel. The calibration gas value begins to flash.22. Use the AIR ▲ YES and RANGE ▼ SHIFT but

Strona 160

232 • Infrared CO2 Calibration EAGLE 2 Operator’s Manualchanges when the unit is turned off and will load the previous set of auto calibration values

Strona 161

EAGLE 2 Operator’s Manual Infrared CO2 Calibration • 23332. If all channels passed calibration, the following screen sequence occurs.If any of the sen

Strona 162

234 • Infrared CO2 Calibration EAGLE 2 Operator’s Manual35. If you want to change the infrared CO2 channel’s calibration gas value, follow Step 17 -

Strona 163

EAGLE 2 Operator’s Manual Infrared CO2 Calibration • 235adjustment for the infrared CO2 channel to the programmed value.40. If the infrared CO2 channe

Strona 164

236 • Maintenance EAGLE 2 Operator’s ManualMaintenanceThis section includes a procedure for replacing an infrared CO2 sensor. Replacing the IR CO2 Sen

Strona 165

EAGLE 2 Operator’s Manual Parts List • 237properly, it may be damaged when the top case is re-installed.16. Make sure that the top case gasket is full

Strona 166

238 • Overview EAGLE 2 Operator’s ManualAppendix H: Infrared Methane SensorOverviewThis appendix describes the EAGLE 2’s infrared methane sensors and

Strona 167

EAGLE 2 Operator’s Manual Description • 239• propane• tolueneThe infrared methane sensor is known to not or to poorly respond to the following combust

Strona 168

EAGLE 2 Operator’s Manual Standard Accessories • 15the O-ring compression force.Charcoal FilterThe charcoal filter is located in the front of the flow

Strona 169

240 • Start Up and Normal Operation EAGLE 2 Operator’s Manualother two sensor chamber locations depending on the particular version of the EAGLE 2. So

Strona 170 - Maintenance

EAGLE 2 Operator’s Manual Infrared Methane Calibration • 241sensor has its own protective shut off, exposure to high levels of combustible gas can sti

Strona 171

242 • Infrared Methane Calibration EAGLE 2 Operator’s Manualmaximum accuracy in the %vol range, the sensor may be calibrated to 50 %vol.To use the aut

Strona 172 - Lamp Window

EAGLE 2 Operator’s Manual Infrared Methane Calibration • 2437. Press and release the POWER ENTER RESET button. The following screen appears.8. Press a

Strona 173

244 • Infrared Methane Calibration EAGLE 2 Operator’s Manualthe probe are tightened firmly together to avoid leaks that can affect the calibration. Se

Strona 174

EAGLE 2 Operator’s Manual Infrared Methane Calibration • 24515. To adjust the values on the screen, press and hold the RANGE ▼ SHIFT button, then pres

Strona 175

246 • Infrared Methane Calibration EAGLE 2 Operator’s Manual22. Press and release the POWER ENTER RESET button. The following screen appears.23. If yo

Strona 176 - Figure 31: Using Removal Tool

EAGLE 2 Operator’s Manual Infrared Methane Calibration • 24727. Press and release the POWER ENTER RESET button to set the span adjustment for each cha

Strona 177

248 • Infrared Methane Calibration EAGLE 2 Operator’s Manual31. If you want to change the infrared methane channel’s calibration gas value, follow Ste

Strona 178 - Replacing the Electrode Stack

EAGLE 2 Operator’s Manual Infrared Methane Calibration • 24936. If the infrared methane channel passed calibration, the following screen sequence occu

Strona 179 - Figure 33: Using Removal Tool

16 • Optional Accessories EAGLE 2 Operator’s ManualCAUTION: Sample hose lengths of more than 125 feet are not recommended for the EAGLE 2 because of f

Strona 180

250 • Maintenance EAGLE 2 Operator’s ManualMaintenanceThis section includes a procedure to replace an infrared methane sensor.Replacing the IR Methane

Strona 181 - Replacing the PID Sensor

EAGLE 2 Operator’s Manual Parts List • 25116. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on

Strona 182 - EAGLE 2 S

252 • Overview EAGLE 2 Operator’s ManualAppendix I: Infrared Hydrocarbon SensorOverviewThis appendix describes the EAGLE 2’s infrared hydrocarbon sens

Strona 183

EAGLE 2 Operator’s Manual Description • 253 Figure 41: Typical Infrared Hydrocarbon Sensor LocationInfrared Hydrocarbon SensorThe infrared hydrocarbon

Strona 184

254 • Start Up and Normal Operation EAGLE 2 Operator’s ManualStart Up and Normal OperationFor instructions to startup and use an EAGLE 2 that includes

Strona 185 - 2000 ppm IBL

EAGLE 2 Operator’s Manual Infrared Hydrocarbon Calibration • 255If set to DISABLED, the gas reading for the catalytic LEL channel will be replaced by

Strona 186

256 • Infrared Hydrocarbon Calibration EAGLE 2 Operator’s Manual5. The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION.6. Mo

Strona 187

EAGLE 2 Operator’s Manual Infrared Hydrocarbon Calibration • 2579. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds,

Strona 188

258 • Infrared Hydrocarbon Calibration EAGLE 2 Operator’s Manual13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES butt

Strona 189

EAGLE 2 Operator’s Manual Infrared Hydrocarbon Calibration • 259channel. The calibration gas value begins to flash.18. Use the AIR ▲ YES and RANGE ▼ S

Strona 190 - ESM-01 Sensor

EAGLE 2 Operator’s Manual Optional Accessories • 17Rechargeable Ni-MH BatteriesRechargeable Ni-MH batteries are available for the EAGLE 2. A fully cha

Strona 191 - ESM-01 Calibration

260 • Infrared Hydrocarbon Calibration EAGLE 2 Operator’s Manualchanges when the unit is turned off and will load the previous set of auto calibration

Strona 192 - NORMAL OPERATION

EAGLE 2 Operator’s Manual Infrared Hydrocarbon Calibration • 26128. If all channels passed calibration, the following screen sequence occurs.If any of

Strona 193

262 • Infrared Hydrocarbon Calibration EAGLE 2 Operator’s Manual31. If you want to change the infrared hydrocarbon channel’s calibration gas value, fo

Strona 194

EAGLE 2 Operator’s Manual Infrared Hydrocarbon Calibration • 26336. If the infrared hydrocarbon channel passed calibration, the following screen seque

Strona 195

264 • Maintenance EAGLE 2 Operator’s ManualMaintenanceThis section includes a procedure to replace an infrared hydrocarbon sensor.Replacing the IR Hyd

Strona 196

EAGLE 2 Operator’s Manual Parts List • 26516. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on

Strona 197

266 • Overview EAGLE 2 Operator’s ManualAppendix J: Methane Elimination ModeOverviewFor applications where methane is an interfering gas, you can set

Strona 198

EAGLE 2 Operator’s Manual Monitoring in Methane Elimination Mode • 2676. You will note that if methane elimination is enabled, the Catalytic Sensor Re

Strona 199

268 • Monitoring in Methane Elimination Mode EAGLE 2 Operator’s ManualWARNING: The EAGLE 2’s alarms are initiated by the display reading, not the fact

Strona 200

EAGLE 2 Operator’s Manual Calibration • 269indicating that the unit is measuring in full response mode.15. Allow 2 minutes for the combustible sensor

Strona 201

18 • Optional Accessories EAGLE 2 Operator’s ManualDC ChargerAn optional DC powered charger is available with a vehicle plug 12 VDC adapter. It uses t

Strona 202

270 • Overview EAGLE 2 Operator’s ManualAppendix K: Using the EAGLE 2 in Bar Hole ModeOverviewThis chapter explains how to operate the EAGLE 2 in Bar

Strona 203

EAGLE 2 Operator’s Manual Start Up, Bar Hole Mode • 271NORMAL MODE and LEAK CHECK MODE will be the only choices displayed. In this case, see “Turning

Strona 204

272 • Start Up, Bar Hole Mode EAGLE 2 Operator’s ManualNOTE: If no button is pressed for 20 seconds, the unit will proceed into whichever mode has the

Strona 205

EAGLE 2 Operator’s Manual Start Up, Bar Hole Mode • 273• If the unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM TO CAL, then the fo

Strona 206

274 • Start Up, Bar Hole Mode EAGLE 2 Operator’s ManualNOTIFICATION ONLY, then the following alert screen displays and the buzzer sounds in a double p

Strona 207

EAGLE 2 Operator’s Manual Start Up, Bar Hole Mode • 275If one of the combustible sensors, LEL or TC, fails, it is not possible to enter Bar Hole Mode.

Strona 208

276 • Bar Hole Testing EAGLE 2 Operator’s Manual3. Press and hold the AIR ▲ YES button. The display will indicate that a demand zero is taking place a

Strona 209

EAGLE 2 Operator’s Manual Bar Hole Testing • 2773. Insert the probe into the bar hole and press and release the ENTER button. The pump will turn on, t

Strona 210 - and HCN Sensors

278 • Bar Hole Testing EAGLE 2 Operator’s Manual• Press and release the DISPLAY ADJUST NO button. The display will now indicate “FRESH AIR PURGE . . .

Strona 211

EAGLE 2 Operator’s Manual Overview • 279Appendix L: Using the EAGLE 2 in Leak Check ModeOverviewThis chapter explains how to operate the EAGLE 2 in Le

Strona 212

EAGLE 2 Operator’s Manual Optional Accessories • 19• 30 inch stainless steel probeFigure 10: 30 Inch Stainless Steel ProbeThis probe is physically the

Strona 213 - Part Number Description

280 • Start Up, Leak Check Mode EAGLE 2 Operator’s Manual4. The LCD will show the following screen for about ten seconds. 5. The Battery Voltage Scree

Strona 214 - Appendix F: TC Sensors

EAGLE 2 Operator’s Manual Start Up, Leak Check Mode • 28110. The gas alarm setpoints are displayed by three screens in sequence: the Low Alarm Screen,

Strona 215 - TC Sub PCB

282 • Start Up, Leak Check Mode EAGLE 2 Operator’s Manualdone with the calibration and exit Calibration Mode, the unit will begin the startup sequence

Strona 216 - Catalytic (LEL) Sensor Screen

EAGLE 2 Operator’s Manual Start Up, Leak Check Mode • 28312. The Date/Time Screen appears for a few seconds.13. If USER/STATION ID is turned on (see “

Strona 217 - CAT (LEL) SENSOR

284 • Leak Testing EAGLE 2 Operator’s Manual15. The EAGLE 2 is now operating in Leak Check Mode. The pump is on and the following screen appears.NOTE:

Strona 218

EAGLE 2 Operator’s Manual Leak Testing • 285• 250 ppm increments from 10,000 ppm to 50,000 ppmAs the gas concentration increases from 0 ppm, the alarm

Strona 219

286 • Leak Testing EAGLE 2 Operator’s ManualTurning Off the EAGLE 2, Leak Check Mode1. Press and hold the POWER button.2. The buzzer will sound and th

Strona 220

EAGLE 2 Operator’s Manual Description • 287Appendix M: Tank Tester ModelThe EAGLE 2 Tank Tester model is intended for checking tanks or vessels that m

Strona 221

288 • Description EAGLE 2 Operator’s ManualFloat Probe AssemblyThe float probe assembly helps prevent liquid from being drawn into the EAGLE 2. The fl

Strona 222

EAGLE 2 Operator’s Manual Start Up and Operation • 289assembly and must be installed on the EAGLE 2 inlet fitting before installing the float probe as

Strona 223

EAGLE 2 Operator’s Manual Table of Contents Table of ContentsChapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 224

20 • Optional Accessories EAGLE 2 Operator’s Manual• 10 inch probe with dust filterFigure 13: 10 Inch Probe With Dust FilterThis probe is designed for

Strona 225

290 • Start Up and Operation EAGLE 2 Operator’s Manual4. The Battery Voltage Screen appears for a few seconds.5. The Inert Mode Selection Screen will

Strona 226

EAGLE 2 Operator’s Manual Alarms • 291If the instrument is sampling a fresh air environment, the oxygen channel will be in alarm. For more information

Strona 227

292 • Parts List EAGLE 2 Operator’s ManualParts ListTable 35 lists part numbers for replacement parts and accessories of the EAGLE 2’s Tank Tester mo

Strona 228

EAGLE 2 Operator’s Manual Description • 293Appendix N: Using the EAGLE 2 in Inert ModeInert Mode is used to measure the combustible gas and/or oxygen

Strona 229

294 • Alarms EAGLE 2 Operator’s Manual3. The LCD will show the following screen for a few seconds with the “S” in the lower right corner indicating th

Strona 230

EAGLE 2 Operator’s Manual Alarms • 2958. Move the cursor next to the oxygen channel. Press and release POWER ENTER RESET.9. A mode select screen will

Strona 231

296 • Start Up EAGLE 2 Operator’s Manual13. When you are done making changes, use RANGE ▼ SHIFT to move the cursor next to END.14. Press and release P

Strona 232

EAGLE 2 Operator’s Manual Start Up • 297WARNING: The EAGLE 2 is not a gas monitoring device until you select a mode and enter it. 5. If you select Nor

Strona 233

298 • Operation EAGLE 2 Operator’s ManualOperationSee “Measuring Mode, Normal Operation” on page 29 for operating instructions keeping in mind that th

Strona 234 - Infrared Sub PCB

EAGLE 2 Operator’s Manual Warranty • 299WarrantyRKI Instruments, Inc. warrants the EAGLE 2 sold by us to be free from defects in materials, workmanshi

Strona 235 - End Caps

EAGLE 2 Operator’s Manual Optional Accessories • 21External Dilution FittingsTwo external dilution fittings are available for the EAGLE 2, a 1:1 dilut

Strona 236 - 0-60 %vol 0.0 %vol

22 • Overview EAGLE 2 Operator’s ManualChapter 3: OperationOverviewThis chapter explains how to use the EAGLE 2 to perform confined space entry monito

Strona 237 - Calibration

EAGLE 2 Operator’s Manual Start Up • 234. The LCD will show the following screen for about ten seconds. 5. The Battery Voltage Screen appears for a fe

Strona 238 - 228 • Infrared CO

24 • Start Up EAGLE 2 Operator’s Manualfrom the last time the EAGLE 2 was used. See “Time in Operation Screen” on page 51 for more information about h

Strona 239 - Calibration • 229

EAGLE 2 Operator’s Manual Start Up • 25• If the unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM TO CAL, then the following screen d

Strona 240 - 230 • Infrared CO

26 • Start Up EAGLE 2 Operator’s Manualwill again appear in the startup sequence.• If the unit is due for calibration and CAL PAST DUE ACT is set to N

Strona 241 - Calibration • 231

EAGLE 2 Operator’s Manual Start Up • 27If you wish to continue, press and release the POWER ENTER RESET button to acknowledge the failure. The gas rea

Strona 242 - 232 • Infrared CO

28 • Using the Battery Charger for Continuous Operation EAGLE 2 Operator’s Man-Using the Battery Charger for Continuous OperationThe battery charger c

Strona 243 - Calibration • 233

EAGLE 2 Operator’s Manual Measuring Mode, Normal Operation • 297. Insert the module’s round plug into the EAGLE 2’s charging jack as shown in Figure 1

Strona 244 - Sensor Calibration Times

Table of Contents EAGLE 2 Operator’s ManualMeasuring Mode, Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Monitoring an

Strona 245 - Calibration • 235

30 • Measuring Mode, Normal Operation EAGLE 2 Operator’s ManualMonitoring an Area1. Start up the EAGLE 2 as described above in “Start Up” on page 22.

Strona 246

EAGLE 2 Operator’s Manual Measuring Mode, Normal Operation • 31The chart below illustrates how response time is affected by the sample hose length.Usi

Strona 247

32 • Measuring Mode, Normal Operation EAGLE 2 Operator’s Manual• The EAGLE 2 provides the catalytic combustible sensor with some protection against ex

Strona 248

EAGLE 2 Operator’s Manual Measuring Mode, Normal Operation • 33CAUTION: Any rapid increase in the combustible gas reading on the catalytic combustible

Strona 249

34 • Measuring Mode, Normal Operation EAGLE 2 Operator’s ManualIf the Catalytic Units menu item in Setup Mode is set to CHANGE OK, the standard factor

Strona 250

EAGLE 2 Operator’s Manual Measuring Mode, Alarms • 35The %LEL and %volume unit readings will still increase up to 100% LEL and 15 %volume respectively

Strona 251

36 • Measuring Mode, Alarms EAGLE 2 Operator’s ManualAlarm IndicationsThe EAGLE 2 will sound an alarm and flash the LED arrays when one of the target

Strona 252

EAGLE 2 Operator’s Manual Measuring Mode, Alarms • 37Resetting and Silencing AlarmsYou can set the EAGLE 2’s gas alarms as latching or self-resetting

Strona 253

38 • Measuring Mode, Alarms EAGLE 2 Operator’s Manualare the factory settings. When the EAGLE 2 goes into gas alarm, press and release any button to s

Strona 254

EAGLE 2 Operator’s Manual Measuring Mode, Alarms • 39Responding to Over Range AlarmsWARNING: An over range condition may indicate an extreme combustib

Strona 255

EAGLE 2 Operator’s Manual Table of Contents Chapter 5: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70Overview .

Strona 256

40 • Display Mode EAGLE 2 Operator’s Manual3. If the sensor failure continues, replace the sensor as described in “Replacing a Sensor” on page 80.4. I

Strona 257

EAGLE 2 Operator’s Manual Display Mode • 41NO button.• To enter an item when a question is asked, press and release the AIR ▲ YES button.• To change a

Strona 258

42 • Display Mode EAGLE 2 Operator’s ManualBattery Voltage ScreenThe Battery Voltage Screen displays the minimum operating voltage and the current bat

Strona 259

EAGLE 2 Operator’s Manual Display Mode • 43If you do not want to change the setting, press and release the DISPLAY ADJUST NO button and the unit will

Strona 260

44 • Display Mode EAGLE 2 Operator’s Manualcatalytic combustible channel.You can select from a list of gases whose response relative to the configured

Strona 261

EAGLE 2 Operator’s Manual Display Mode • 45The following is the complete list of factory defined gases.2. Use the AIR ▲ YES or RANGE ▼ SHIFT buttons t

Strona 262

46 • Display Mode EAGLE 2 Operator’s ManualSTEL ScreenThe STEL Screen displays the short term exposure limit (STEL) readings for H2S and CO only. The

Strona 263 - Infrared Hydrocarbon Sensor

EAGLE 2 Operator’s Manual Display Mode • 47AIR ▲ YES button. The following screen appears showing Channel 1 alarm points.2. If the channel description

Strona 264

48 • Display Mode EAGLE 2 Operator’s Manual4. Press and release the AIR ▲ YES button to scroll through screens that display the rest of the active cha

Strona 265

EAGLE 2 Operator’s Manual Display Mode • 49user IDs stored in the EAGLE 2, see the Eagle 2 Maintenance Data Loader Program Operator’s Manual.To select

Strona 266

Table of Contents EAGLE 2 Operator’s ManualUpdating the Alarm Point Settings . . . . . . . . . . . . . . . . . . . . . . . . . 115Updating the Alarm

Strona 267

50 • Display Mode EAGLE 2 Operator’s ManualSelect Station ID ScreenThis screen displays only if USER/STATION ID in the Setup Mode menu is set to ON (s

Strona 268

EAGLE 2 Operator’s Manual Display Mode • 51screen if the cursor is moved up from the first station ID or to the next screen if the cursor is moved dow

Strona 269

52 • Display Mode EAGLE 2 Operator’s ManualDate/Time ScreenThe Date/Time Screen displays the current date and time.Data Logging ScreenCAUTION: Once yo

Strona 270

EAGLE 2 Operator’s Manual Data Logging • 533. If you want to clear the data logger memory, press and release the AIR ▲ YES button. The following scree

Strona 271

54 • Data Logging EAGLE 2 Operator’s ManualData Logger Management Program and a computer with an infrared port or a USB port that runs one of the foll

Strona 272

EAGLE 2 Operator’s Manual Overview • 55Chapter 4: Calibration ModeOverviewThis section describes the EAGLE 2 in Calibration Mode. In Calibration Mode,

Strona 273

56 • Calibration Supplies and Equipment EAGLE 2 Operator’s ManualCalibration Supplies and EquipmentTo calibrate the EAGLE 2, you will need:• Known cal

Strona 274

EAGLE 2 Operator’s Manual Entering Calibration Mode • 57cylinder. This chapter also includes instructions for calibrating one channel at a time using

Strona 275

58 • Calibrating Using the Auto Calibration Method EAGLE 2 Operator’s ManualCalibrating Using the Auto Calibration MethodThis method allows you to cal

Strona 276 - DISABLED

EAGLE 2 Operator’s Manual Calibrating Using the Auto Calibration Method • 594. The EAGLE 2 will indicate that it is adjusting the zero reading for a f

Strona 277

EAGLE 2 Operator’s Manual Table of Contents PID Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137Star

Strona 278

60 • Calibrating Using the Auto Calibration Method EAGLE 2 Operator’s Manual4. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR

Strona 279 - Calibration

EAGLE 2 Operator’s Manual Calibrating Using the Auto Calibration Method • 61channel. The calibration gas value begins to flash.9. Use the AIR ▲ YES an

Strona 280 - Start Up, Bar Hole Mode

62 • Calibrating Using the Auto Calibration Method EAGLE 2 Operator’s Manualchanges when the unit is turned off and will load the previous set of auto

Strona 281

EAGLE 2 Operator’s Manual Calibrating Using the Auto Calibration Method • 6319. If all channels passed calibration, the following screen sequence occu

Strona 282

64 • Calibrating Using the Single Calibration Method EAGLE 2 Operator’s ManualCalibrating Using the Single Calibration MethodSingle Calibration allows

Strona 283

EAGLE 2 Operator’s Manual Calibrating Using the Single Calibration Method • 654. The EAGLE 2 will indicate that it is adjusting the zero reading for a

Strona 284

66 • Calibrating Using the Single Calibration Method EAGLE 2 Operator’s Manual4. Move the cursor next to the SINGLE CALIBRATION menu item by using the

Strona 285 - ADJUST: PURGE

EAGLE 2 Operator’s Manual Calibrating Using the Single Calibration Method • 677. Press and release the POWER ENTER RESET button to proceed to the Sing

Strona 286 - Bar Hole Testing

68 • Calibrating Using the Single Calibration Method EAGLE 2 Operator’s Manualresponse level to the applied gas. This calculated range is independent

Strona 287

EAGLE 2 Operator’s Manual Calibrating Using the Single Calibration Method • 69CAUTION: When calibrating the oxygen sensor, verify the concentration of

Strona 288

Table of Contents EAGLE 2 Operator’s ManualTC Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 289 - Start Up, Leak Check Mode

70 • Overview EAGLE 2 Operator’s ManualChapter 5: MaintenanceOverviewThis chapter describes troubleshooting procedures for the EAGLE 2. It also includ

Strona 290

EAGLE 2 Operator’s Manual Troubleshooting • 71• The unit indicates flow failure and does not recover when POWER ENTER RESET is pressed and released.•

Strona 291

72 • Replacing or Recharging the Batteries EAGLE 2 Operator’s ManualReplacing or Recharging the BatteriesWARNING: To prevent ignition of a hazardous a

Strona 292

EAGLE 2 Operator’s Manual Replacing or Recharging the Batteries • 73Figure 16: Removing the Battery Case3. Pull the battery case away from the bottom

Strona 293

74 • Replacing or Recharging the Batteries EAGLE 2 Operator’s ManualRecharging the Ni-MH BatteriesCAUTION: Use with Ni-MH battery p/n 49-1330RK. Charg

Strona 294 - CH4 0 ppm

EAGLE 2 Operator’s ManualReplacing the Hydrophobic Probe’s Particle Filter and Hydro-6. While the batteries are charging, the green indicator LED will

Strona 295 - Locating a Leak

76 • Replacing the Hydrophobic Probe’s Particle Filter and Hydrophobic FilterFigure 19: Replacing the Particle Filter and Hydrophobic Filter Disk2. Re

Strona 296

EAGLE 2 Operator’s Manual Replacing the Hydrophobic Filter • 777. Place the new filter disk flat on top of the particle filter. Make sure it is center

Strona 297 - Appendix M: Tank Tester Model

78 • Replacing the Hydrophobic Filter EAGLE 2 Operator’s Manual6. Locate the hydrophobic filter. It is over the oxygen sensor. Note which side of the

Strona 298

EAGLE 2 Operator’s Manual Replacing the Charcoal Filter • 79Replacing the Charcoal Filter1. Verify that the EAGLE 2 is off.2. Place the EAGLE 2 upside

Strona 299 - CONNECT FLOAT PROBE

EAGLE 2 Operator’s Manual Table of Contents Infrared Hydrocarbon Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 253Infrared Sub PCB.

Strona 300

80 • Checking the Combustible Gas Sensor’s ConditionEAGLE 2 Operator’s Manualthe flow chamber and push it in until it bottoms out.9. Insert the charco

Strona 301

EAGLE 2 Operator’s Manual Replacing a Sensor • 815. Lay the top case down next to the bottom case to allow access to the flow system.6. Locate the sen

Strona 302

82 • Replacing a Sensor EAGLE 2 Operator’s Manual• If the sensor stayed in the flow chamber, grasp the sensor and pull it out of the flow chamber.8. T

Strona 303

EAGLE 2 Operator’s Manual Replacing a Sensor • 83• Insert the LEL sensor into the LEL sensor chamber in the flow chamber.• Line up the holes in the LE

Strona 304

84 • General Parts List EAGLE 2 Operator’s ManualGeneral Parts ListTable 11 lists part numbers for the EAGLE 2’s replacement parts and accessories.Ta

Strona 305

EAGLE 2 Operator’s Manual General Parts List • 8565-2005RK Carbon monoxide (CO) sensor71-0154RK Operator’s Manual, EAGLE 2 (this document)71-0170RK Op

Strona 306 - Start Up

86 • Overview EAGLE 2 Operator’s ManualAppendix A: Calibrating with a Sample BagOverviewThe EAGLE 2 can be calibrated with a gas bag calibration kit i

Strona 307

EAGLE 2 Operator’s Manual Entering Calibration Mode • 87To calibrate the EAGLE 2, you will need:• Known calibrating samples of the gases being detecte

Strona 308 - NORMAL MODE

88 • Calibrating Using the Auto Calibration Method EAGLE 2 Operator’s Manual3. If the unit prompts you for the password, enter it by using the AIR ▲ Y

Strona 309 - Warranty

EAGLE 2 Operator’s Manual Calibrating Using the Auto Calibration Method • 892. Press and release the POWER ENTER RESET button. The following screen ap

Komentarze do niniejszej Instrukcji

Brak uwag