1. Turning the GX-2009 On
a.PressandbrieyholddownthePOWERMODEbutton.The
backlightwillturnonandactivateallLCDsegments.Release
thebuttonwhenyouhearabeep.
i.IfCalC-LIMITisdisplayed,unitispastdueforcalibration.Pressand
releasethePOWERbuttontoconrmandcontinueuse.
ii. IfFAILC-LIMITisdisplayed,unitispastdueforcalibrationandcannotbe
usedwithoutcalibrating.
b.Theinstrumentwillgothroughabriefwarmupshowingdate,
numberofdaystonextcalibration,timeofday,batterypower,
alarmlatchsetting,fullscalevaluesforallchannels,warning
alarmsettingandhighalarmsettingforallchannels,STEL
alarmsetpointforH2SandCOandTWAalarmsetpointfor
H2SandCOchannels.
c.Ifthebatterylevelisloworthebatteryiconisashingandthe
unit’sbuzzersounds,rechargetheGX-2009beforeuse.
2. Measuring Mode Screen
a.Afterwarm-uptheGX-2009willdisplaythecurrentO2,H2S,
COand%LELgasreadings.
b.Ifthereadingsarenotdisplayingfreshairvalues,itis
necessarytoperformaFreshAiradjustment.
3. Performing a Fresh Air Adjustment
a.Findafresh-airenvironmentfreeoftoxicorcombustible
gasesandofnormaloxygencontent(20.9%).
b.PressandholdtheAIRbuttonforaboutthreeseconds.The
displaywillindicateAirCALHOLDAIR.Continuetoholduntil
ADJandRELEASEaredisplayed.ReleasetheAIRbutton.
TheGX-2009willsettofreshairvalues,0%LELCH4,20.9%
O2,0.0ppmH2Sand0ppmCO.
4. Performing a Breath Test
a.PerformingabreathtestquicklyteststheGX-2009oxygen
sensor,audible,visualandvibratoryalarmsforproper
operation.Performingabreathtestisnotasubstitutefor
calibrationorperformingabumptestonallsensors.
b.CuphandsoverO2sensoropeningandgentlyexhaleover
thesensor.
c.Theoxygensensoroutputwilldropbelowthealarmpointof
19.5%andactivatealarms.
d.Ifalarmsaresettolatching,pressthePOWERMODEbutton
toresetalarmsonceoxygenreadingisabove19.5%.
i. Note:YoumayalsoseeslightresponseontheCOgassensorwhen
performingabreathtest.
Quick Reference Guide For Model GX-2009
Turn on and adjust the GX-2009 gas monitor in a known fresh air area.
Przeglądanie stron 0
1 2

Podsumowanie treści

Strona 1 - 4. Performing a Breath Test

1. Turning the GX-2009 Ona.PressandbrieyholddownthePOWERMODEbutton.ThebacklightwillturnonandactivateallLCDsegments.Releasethe

Strona 2 - 081003/1000

Quick Reference Guide For Model GX-2009PN71-0160RKRevP1RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 Phone(510)441-5656•(800)7

Komentarze do niniejszej Instrukcji

Brak uwag