www.rkiinstruments.com35-3001A-09Oxygen/Hydrogen SulfideSample-Draw DetectorOperator’s ManualPart Number: 71-0311RKRevision: P3Released: 9/17/14

Strona 2

6 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorOxygen AmplifierThe amplifier converts the electrical output from the detector to a 4 to

Strona 3 - Product Warranty

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 7Hydrogen Sulfide Sensor/H2S AmplifierThe hydrogen sulfide sensor is a plug-in sensor tha

Strona 4 - Table of Contents

8 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorInstallationThis section describes procedures to mount the sample-draw detector in the mo

Strona 5 - Specifications

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 92. Close and latch the housing door.NOTE: The sample-draw detector is shipped with the m

Strona 6 - Description

10 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detectorring, if it came out, in the orientation shown in Figure 6. 4. Screw the inlet fitting t

Strona 7 - Internal Components

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 11Installing the Exhaust Line1. Loosen the nut on the exhaust fitting until 3 threads are

Strona 8

12 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector8. Connect the wires to the applicable detector/transmitter terminal strip at the contro

Strona 9 - Pump Reset Switch

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 139. If shielded cable is used, connect the cable’s drain wire to an available chassis (e

Strona 10 - Test Points

14 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector3. Set a voltmeter to measure in the millivolt (mV) range.4. Plug the voltmeter leads in

Strona 11 - 100-500 mV Range

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 15Preparing for the response testNOTE: This procedure describes the RKI calibration kit t

Strona 12 - Installation

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of

Strona 13 - Back Ferrule

16 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector12. Reconnect the sample tubing to the inlet fitting.Quarterly CalibrationCalibrate the

Strona 14

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 17• The sample-draw detector is malfunctioning.Recommended action1. Verify that the calib

Strona 15

18 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detectorand sensor off of the flow block.There is a gasket at the bottom of the flow block. Be s

Strona 16 - White (+)

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 19NOTE: When the sample-draw detector is first powered up with a new amplifier, the initi

Strona 17 - Start Up

20 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorReplacing the Hydrophobic Filter1. Open the housing door of the sample-draw detector.2.

Strona 18 - Maintenance

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 21Calibration FrequencyAlthough there is no particular calibration frequency that is corr

Strona 19 - Performing the response test

22 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detectorsection, Setting the Zero Reading.5. Disconnect the sample tubing from the inlet fitting

Strona 20 - Troubleshooting

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 232. Screw the regulator into a zero air calibration cylinder.3. Set a voltmeter to measu

Strona 21 - Recommended action

24 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorParts ListTable 2 lists replacement parts and accessories for the sample-draw detector.T

Strona 22

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free fro

Strona 23

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 24

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 1OverviewThis operator’s manual describes the 35-3001A-09 oxygen/hydrogen sulfide sample-

Strona 25 - Calibration, Oxygen Sensor

2 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorDescriptionThis section describes the components of the oxygen/hydrogen sulfide sample-dr

Strona 26 - S Sensor

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 3installation where general purpose equipment is in use.The housing door is hinged on the

Strona 27 - Returning to Normal Operation

4 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorDetector/Amp Terminal StripThe detector/amp terminal strip is the upper twelve-point term

Strona 28 - Parts List

35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 5Sensor Flow Control ValveThe sensor flow control valve is mounted to the flowmeter circu