www.rkiinstruments.com35-3001A-09Oxygen/Hydrogen SulfideSample-Draw DetectorOperator’s ManualPart Number: 71-0311RKRevision: P3Released: 9/17/14
6 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorOxygen AmplifierThe amplifier converts the electrical output from the detector to a 4 to
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 7Hydrogen Sulfide Sensor/H2S AmplifierThe hydrogen sulfide sensor is a plug-in sensor tha
8 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorInstallationThis section describes procedures to mount the sample-draw detector in the mo
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 92. Close and latch the housing door.NOTE: The sample-draw detector is shipped with the m
10 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detectorring, if it came out, in the orientation shown in Figure 6. 4. Screw the inlet fitting t
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 11Installing the Exhaust Line1. Loosen the nut on the exhaust fitting until 3 threads are
12 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector8. Connect the wires to the applicable detector/transmitter terminal strip at the contro
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 139. If shielded cable is used, connect the cable’s drain wire to an available chassis (e
14 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector3. Set a voltmeter to measure in the millivolt (mV) range.4. Plug the voltmeter leads in
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 15Preparing for the response testNOTE: This procedure describes the RKI calibration kit t
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of
16 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector12. Reconnect the sample tubing to the inlet fitting.Quarterly CalibrationCalibrate the
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 17• The sample-draw detector is malfunctioning.Recommended action1. Verify that the calib
18 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detectorand sensor off of the flow block.There is a gasket at the bottom of the flow block. Be s
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 19NOTE: When the sample-draw detector is first powered up with a new amplifier, the initi
20 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorReplacing the Hydrophobic Filter1. Open the housing door of the sample-draw detector.2.
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 21Calibration FrequencyAlthough there is no particular calibration frequency that is corr
22 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detectorsection, Setting the Zero Reading.5. Disconnect the sample tubing from the inlet fitting
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 232. Screw the regulator into a zero air calibration cylinder.3. Set a voltmeter to measu
24 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorParts ListTable 2 lists replacement parts and accessories for the sample-draw detector.T
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free fro
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 1OverviewThis operator’s manual describes the 35-3001A-09 oxygen/hydrogen sulfide sample-
2 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorDescriptionThis section describes the components of the oxygen/hydrogen sulfide sample-dr
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 3installation where general purpose equipment is in use.The housing door is hinged on the
4 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw DetectorDetector/Amp Terminal StripThe detector/amp terminal strip is the upper twelve-point term
35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 5Sensor Flow Control ValveThe sensor flow control valve is mounted to the flowmeter circu