Rki-instruments 01 Series Quik Start Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Czujniki Rki-instruments 01 Series Quik Start. RKI Instruments 01 Series Quik Start User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 2
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
Quick Reference Guide For 01-Series
Note: Turn on and adjust the 01-Series gas monitor in a known fresh air area.
1. Press and hold POWER/MODE button for 3
seconds to turn instrument ON
2. Warm-up sequence
a. All segments of LCD turn on.
b. LCD back light turns on.
c. Buzzer sounds and vibrator activates.
d. Alarm lights flash.
e. Instrument shows target gas and battery voltage.
f. Buzzer will sound after battery voltage is displayed.
g. Instrument is now in normal operating mode.
Note: If instrument gives a low battery or dead battery alarm,
replace batteries or recharge (GP-01 only) before use.
3. Performing Fresh Air Adjustment
a. Press and hold the AIR button. The LCD displays “hold”.
b. Release the AIR button when LCD displays “Adj”.
c. The instrument will set reading to 0 ppm (H2S/CO), 0% LEL
or 20.9% Vol. oxygen depending on the instrument.
4. Displaying PEAK, STEL and TWA (CO and H2S
versions)
a. Press and release the POWER/MODE button to display the
PEAK gas reading.
b. Press and release the POWER/MODE button again to
display STEL reading.
c. Press and release the POWER/MODE button again to
display the TWA reading.
d. Press and release POWER/MODE button again to return to
normal measuring mode.
Note: If no button is pressed for 20 seconds, the instrument will
automatically return to measuring mode.
Przeglądanie stron 0
1 2

Podsumowanie treści

Strona 1

Quick Reference Guide For 01-SeriesNote: Turn on and adjust the 01-Series gas monitor in a known fresh air area.1. Press and hold POWER/MODE button f

Strona 2 - 6. Changing the Alarm Points

Quick Reference Guide For 01-SeriesPN 72-0141RK Rev 0RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 Phone (510) 441-5656 • (800) 754-51

Komentarze do niniejszej Instrukcji

Brak uwag