Setting Commands 45
is received as new if the factory settings for them are not how
you want them.
The Factory Settings Window that is accessible with the Detail
Settings button is shown below with the factory settings for
each of these instrument parameters. If any of these settings is
not set as you want, see the instructions in “Updating an
Instrument’s Detail Settings” on page 50 to change them. The
Detail Settings are described below.
Alarm Pattern
With Alarm pattern set to Latch (factory setting), the GX-2009
remains in alarm condition until the alarm condition passes and
the POWER MODE button is pressed and released.
With Alarm pattern set to Reset, the GX-2009 automatically
resets an alarm when the alarm condition passes.
Figure 38: GX-2009 Factory Settings Window