Rki-instruments GX-2003 Instrukcja Użytkownika Strona 58

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 64
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 57
Calibrating A GX-2003 With the Data Logging Software 58
6. Use the selection boxes under the Span column to select the
channels you want to calibrate.
NOTE: If installed, the TC % volume sensor cannot be
calibrated at the same time as the other sensors
because a multi-gas cylinder that includes % volume
methane is not available.
7. Click the Zero button. The software will perform a zero
adjustment setting the oxygen channel to 20.9% and all
other channels to 0.
Figure 39: Set Window with Calibration Frame
Set Auto Cal. values to matchClick boxes to select
channels for calibration
gas cylinder
Przeglądanie stron 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 64

Komentarze do niniejszej Instrukcji

Brak uwag