RI-215A Operator’s Manua l Part Number: 71-0045RKRevision 0Released: 10/3/05
10 • Overview RI-215A Operator’s ManualChapter 3: InstallationOverviewThis chapter describes how to install the RI-215A at the monitoring site. This
RI-215A Operator’s Manual Mounting the RI-215A • 113. Use two screws to secure the base plate to the wall. Figure 2 below shows the location of the s
12 • Wiring the RI-215A RI-215A Operator’s ManualWiring the RI-215AThis section describes procedures to connect power to the RI-215A. This section al
RI-215A Operator’s Manual Wiring the RI-215A • 13Figure 4 below shows how to wire the RI-215A to a controller that accepts 3-wire, 4-20 mA transmitte
14 • Start Up RI-215A Operator’s ManualConnecting the RI-215A to a ControllerPerform the following procedure to connect the RI-215A as a 3-wire 4 - 2
RI-215A Operator’s Manual Start Up • 15Preparing to Set the ZeroConnect the sample tubing from your calibration kit to the hose barb on the fixed flow
16 • Overview RI-215A Operator’s ManualChapter 4: OperationOverviewThis chapter describes Measuring Mode and Alarm Adjust Mode. Under Measuring Mode,
RI-215A Operator’s Manual Measuring Mode • 17Analog OutputThe RI-215A’s analog output, available via the yellow and blue wires of the cable assembly
18 • Alarm Adjust Mode RI-215A Operator’s ManualSo if the alarm relay contacts are wired to turn on a ventilation system when the concentration of CO
RI-215A Operator’s Manual Alarm Adjust Mode • 19Adjusting the Alarm PointTo adjust the alarm point, do the following:1. Press and hold the Up (∆) swi
2 • Warranty RI-215A Operator’s Manual Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in material
20 • Overview RI-215A Operator’s ManualChapter 5: MaintenanceOverviewThis chapter describes maintenance and troubleshooting procedures for the RI-215
RI-215A Operator’s Manual Replacing Components • 21Replacing ComponentsThe section includes procedures to replace the sensor and fuse.Replacing the D
22 • Calibration RI-215A Operator’s ManualCalibrationThis section describes how to calibrate the RI-215A. It includes procedures to prepare for calib
RI-215A Operator’s Manual Parts List • 233. If not already in, insert the calibration tubing into the gas inlet opening at the bottom of the RI-215A
24 • Parts List RI-215A Operator’s ManualAppendix A: Changing the Detection RangeThe RI-215A can be setup for one of three detection ranges: 0 - 2,00
RI-215A Operator’s Manual Table of Contents • 3 Table of Contents Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 • Table of Contents RI-215A Operator’s ManualVerifying the Alarm Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RI-215A Operator’s Manual Overview • 5Chapter 1: IntroductionOverviewThis chapter briefly describes the RI-215A. This chapter also describes the RI-21
6 • Specifications RI-215A Operator’s ManualSpecificationsTable 1 lists specifications for the RI-215A.1 A 0.5A, 250 V fuse is an acceptable replacemen
RI-215A Operator’s Manual Overview • 7Chapter 2: DescriptionOverviewThis section describes the external and internal components of the RI-215A.Extern
8 • Internal Description RI-215A Operator’s ManualOn the bottom of the front housing is a series of slots and a calibration port. The slots allow the
RI-215A Operator’s Manual Internal Description • 9LCDThe LCD is mounted to the printed circuit board. The LCD shows the current gas reading of the RI